Real-time imaging of boron concentration during irradiation · revision history
Revision 1
Accepted · first revision · record
- Title
- Real-time imaging of boron concentration during irradiation
- Plain
- Each boron capture emits a gamma ray of a specific energy, which could in principle be used to map the boron during treatment. A working scanner that does this accurately in a patient does not exist.
- Precise
- Image the B-10 concentration in tissue during neutron irradiation through the 478 keV gamma line from the Li-7 excited state, with 10 percent accuracy in concentration, about 5 mm spatial resolution and a frame time of minutes, against the intense background of 2.2 MeV hydrogen-capture gammas and neutron-induced detector activation. An answer is a demonstrated detector and reconstruction method with these figures in a patient-sized phantom.
- Settled by
- Phantom measurements with known boron distributions in a clinical epithermal beam showing the stated concentration accuracy and resolution.
- Kind
- well-posed
- Answer type
- value
- References
- (empty)
- Literature status
- open
- Status note
- Prompt-gamma tomography prototypes for neutron capture therapy were in development without clinical use.
- Posed since
- (empty)