Beam-quality factors and $W_{\mathrm{air}}$ for ion dosimetry
Plain
Converting a detector reading into dose in a particle beam needs the average energy spent to make one ion pair in air. Its uncertainty, especially for carbon ions, dominates the dose error.
Precise
Reduce the uncertainty in the mean energy per ion pair in air $W_{\mathrm{air}}$ and in the beam-quality correction factor $k_Q$ for carbon and other ions to the level needed for a total dosimetric uncertainty below $1.5\,\mathrm{percent}$. ICRU 90 values are $W_{\mathrm{air}}$ of about $34.44\,\mathrm{eV}$ for protons (about $0.4\,\mathrm{percent}$ standard uncertainty) and $34.71\,\mathrm{eV}$ for carbon ions (about $1.5\,\mathrm{percent}$). An answer is a measured value set with reduced uncertainty.
Settled by
Calorimetric determination of $W_{\mathrm{air}}$ and $k_{Q}$ in clinical ion beams.
Kind
well-posed
Answer type
value
References
(empty)
Literature status
partially-resolved
Status note
IAEA TRS-398 Rev. 1 (2024) adopted the ICRU 90 data; for carbon ions the $W_{\mathrm{air}}$ and $k_{Q}$ terms still keep the total uncertainty near 3 percent.
Posed since
1990