Transcranial focusing without a prior CT scan · revision history
Revision 1
Accepted · first revision · record
- Title
- Transcranial focusing without a prior CT scan
- Plain
- The skull bends and absorbs sound unevenly, smearing the focus inside the head. Correcting for it currently needs an x-ray scan of the skull.
- Precise
- Achieve transcranial focal placement accurate to 1 mm and pressure prediction accurate to 10 percent at 0.2 to 1 MHz without computed tomography of the skull, by inferring the skull acoustic properties (sound speed $2.4\,\mathrm{x}\,10^{5} \text{ to } 3.0\,\mathrm{x}\,10^{5}\,\mathrm{cm}/\mathrm{s}$, density, absorption, shear conversion) from ultrasound or magnetic resonance data. An answer is a method with verified focal position and amplitude in human skulls.
- Settled by
- Hydrophone measurement inside ex vivo human skulls, and in vivo displacement or thermometry confirmation, against the predicted focus.
- Kind
- well-posed
- Answer type
- value
- References
- (empty)
- Literature status
- open
- Status note
- (empty)
- Posed since
- 2000