Optimal energy thresholds for material separation · revision history

Revision 1

Accepted · first revision · record

Title
Optimal energy thresholds for material separation
Plain
The detector sorts photons into a few energy bins. Where to place the bin edges for the best material separation at the lowest dose is unsolved in general.
Precise
Determine the energy threshold set that minimizes the variance of material-decomposition estimates for a given task, tube spectrum, patient thickness and dose, including the detector response with charge sharing and pileup. Compare against the Cramer-Rao bound (the minimum variance any unbiased estimator can reach) achievable with full spectral information. An answer is an optimization rule plus the efficiency loss from using 2 to 8 bins.
Settled by
Derivation of the bound plus phantom measurement confirming the predicted variance for optimized thresholds.
Kind
well-posed
Answer type
value
References
(empty)
Literature status
partially-resolved
Status note
Information-theoretic treatments exist for idealized detectors; the realistic-response case stays open.
Posed since
2010